Please use this identifier to cite or link to this item:
https://repo.btu.kharkov.ua//handle/123456789/494
Title: | Determination of the content of l-ascorbic acid in fruit jelly |
Other Titles: | Визначення вмісту аскорбінової кислоти у фруктовому желе |
Authors: | Gubsky, Sergey Dobrovolska, Olena Aksonova, Olena Evlash, Viktoriya Zheleznyak Zinaida |
Keywords: | ascorbic acid;fruit jelly;galvanostatic coulometry;validation;аскорбінова кислота;фруктове желе;гальваностатична кулонометрія;валідація;аскорбиновая кислота;фруктовое желе;гальваностатическая кулонометрия;валидация |
Issue Date: | 2018 |
Publisher: | Харківський державний університет харчування та торгівлі |
Citation: | Gubsky S., Dobrovolska O., Aksonova O., Evlash V., Zhelezniak Z. Determination of the content of l-ascorbic acid in fruit jelly. Прогресивні техніка та технології харчових виробництв ресторанного господарства і торгівлі. 2018. Вип. 2(28). C. 262-278. |
Abstract: | The article presents the results of the development of a method for the
determination of L-ascorbic acid (AA) in model fruit jellies containing gelatin. In
the determination of AA, a coulometric method with electrically generated iodine as
a titrant in a galvanostatic mode with a current of 2−5 mA was used. As a research
object, model fruit jelly with 2% and 3% gelatin content was used. In addition to
gelatin, the systems contained apple juice concentrated clarified and potassium
sorbate (used as a concerto). The peculiarity of the preliminary preparation of the
samples was the stage, which envisaged the melting of the samples before the
quantitative determination of AA.
In order to assess the suitability of the proposed methodology for the
quantitative determination of AA in fruit jelly, a validation evaluation was
performed on the following parameters: specificity, linearity and the range of an
analytical method, limits of detection and determination, accuracy and
reproducibility. Specificity has been studied on standard aqueous solutions of AA. It
has been shown that under two conditions: 100% of current output and
stoichiometricity of the oxidation reaction with iodine AA, the theoretical slope of
the amount of the electrice curve from the analyte concentration with accuracy up to
0.2% coincides with the experimental one. The following parameters were
determined: the working area of the technique 4.92−127.6 mg/100 g; limits of the
determination 1.62 mg/100 g and quantitation 4.92 mg/100 g. The absence of a
systematic error for standard and fruit jelly solutions was evaluated using the
"introduced-found" method. Error of recovery did not exceed RSD<2.8%.
Correctness and reproducibility were evaluated by the method of varying the
weights at three levels of concentration using three sets at each level. Fischer and
Student's criteria, which were used to compare average values, testified to the
validity of the indicators of "correctness" and "reproducibility". Comparing the
studied parameters, the best results were obtained for fruit jelly with a 2% gelatin
content compared to a jelly with a content of 3%.
The obtained results testify to the possibility of application of the developed
method of determination of ascorbic acid in real food systems. Наведено результати розробки методики визначення аскорбінової кислоти в модельних фруктових желе, що містять желатин. Запропонована методика базується на кулонометричному титруванні електрогенерованим йодом у гальваностатичному режимі розплавлених зразків фруктового желе. Досліджено параметри валідаційної оцінки: специфічність, лінійність, робоча область методики, межі виявлення та визначення, правильність та відтворюваність. Отримані результати свідчать про можливість застосування розробленої методики визначення аскорбінової кислоти в реальних харчових системах. |
URI: | https://repo.btu.kharkov.ua//handle/123456789/494 |
Appears in Collections: | Випуск 2 (28) |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Pt_2018_2_25.pdf | 429.38 kB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.